medicine.kazimiri
Histology Exam 1 · Lecture 1

Microscopy

Cheat sheet: your lecture slides, checked against Junqueira. Histology Exam 1 · PDF of all lectures · printout sheets (front and back PDFs: Basic, Intermediate, Advanced)
Next ›

1 High-yield facts

ResolutionSmallest distance at which two points are still seen as separate; d = 0.61λ / NA. Shorter wavelength or higher NA gives finer resolution.
LimitsEye ~0.2 mm; light microscope ~0.2 µm; TEM ~1–3 nm in tissue sections (theoretical far less).
MagnificationTotal = objective × eyepiece (40× × 10× = 400×). Magnifying past the resolution limit is empty magnification: no new detail.
Immersion oilOil (n ≈ 1.515, like glass) between coverslip and 100× objective raises NA to ~1.4, so resolution improves.
Worked exampleGreen light λ 550 nm, NA 1.4: d = 0.61 × 550 / 1.4 ≈ 240 nm = 0.24 µm.
Units1 mm = 1000 µm; 1 µm = 1000 nm. RBC ~7.5 µm; ribosome ~25 nm (TEM only).
Bright-field partsLight source, condenser (with diaphragm), stage, objective, eyepiece; image from light absorbed by stained tissue.
Working routineFind and focus with the low-power objective first; bring a high-power lens to the coverslip watching from the side, then focus by moving it away.
Electron microscopyElectrons have a far shorter wavelength than light; lenses are electromagnetic coils; vacuum column; image seen on a screen or camera.
TEM contrastHeavy metals (osmium, uranium, lead) scatter electrons: bound areas look dark (electron-dense), others pale (electron-lucent).
FluorescenceFluorochrome absorbs short-wavelength light and emits longer-wavelength light (Stokes shift); fluorescein green, rhodamine red, DAPI blue.
Modern solutionsConfocal and multiphoton optical sectioning, super-resolution (Nobel 2014), cryo-EM (Nobel 2017), digital whole-slide imaging.
Basophilia and acidophiliaBasic dyes (haematoxylin) bind acidic groups, e.g. nucleic acid phosphates; acidic dyes (eosin) bind basic proteins of cytoplasm and collagen.
Labelling methodsImmunohistochemistry (labelled antibody), autoradiography (silver grains over radioactive precursor), in situ hybridisation (labelled probe).

2 Cheat sheet

Basic shows the must-knows, Intermediate everything on your slides, Advanced adds textbook detail (marked +).

Light microscopes

TypePrincipleBest for
Bright-fieldtransmitted light absorbed by stained sectionroutine stained (H&E) sections
Phase contrastconverts differences in refractive index into light and darkliving, unstained cells (cell culture)
Differential interference contrast (Nomarski)polarised beams give pseudo-3D reliefunstained cells, surface relief
Dark-fieldonly light scattered by the specimen enters the objectivebright objects on black: spirochaetes, crystals
Polarisingpolariser + analyser; birefringent structures shinecollagen, skeletal muscle A band, crystals, amyloid (Congo red)
FluorescenceUV/blue excitation; longer-wavelength emission seenfluorochrome-labelled antibodies, DNA dyes
Confocal (laser scanning)pinhole rejects out-of-focus light; point scanningoptical sections and 3D reconstruction of thick tissue
Multiphotontwo long-wavelength photons excite one fluorochromedeep imaging of living tissue

Electron microscopes

TypeImage formed fromSpecimenShows
Transmission (TEM)electrons passing through the sectionultrathin section 40–90 nm, heavy-metal stainedultrastructure: organelles, ribosomes, membranes
Scanning (SEM)secondary electrons from the surfacewhole specimen coated with gold/metal3D surface of cells and tissues
Freeze-fracturemetal replica of fractured frozen tissuecleaved along lipid bilayerinner membrane faces (P and E), intramembrane proteins
Cryo-EMelectrons through frozen samplemolecules in vitreous ice, unstainednear-atomic protein structure (Nobel 2017)
Cryo-electron tomographytilt series reconstructed in 3Dfrozen cells or thin lamellaemolecular architecture in situ

Resolution: what decides it

FactorEffectNote
Wavelength (λ)shorter λ gives finer dblue/UV better than red; electrons best
Numerical aperture (NA = n sin α)higher NA gives finer ddry objective ≤0.95; oil ~1.4
Refractive index (n) of mediumoil raises nair 1.0, water 1.33, oil 1.515
Condenser settingmust match objective NAclosing diaphragm adds contrast, loses resolution
Magnificationdoes not change dbeyond ~1000× NA is empty magnification

Scale: what each instrument can see

ObjectSizeSeen with
Oocyte~100 µmnaked eye (just) / LM
Red blood cell~7.5 µmLM
Bacterium~1–2 µmLM
Mitochondrion~0.5–1 µm wideLM (barely, special stains); detail by TEM
Cell membrane~7.5–10 nmTEM
Ribosome~25 nmTEM
Microtubule~25 nmTEM
Actin filament~6–7 nmTEM
Protein molecule~2–10 nmcryo-EM

Modern and super-resolution methods

MethodHow it beats the limit or addsResolution / use
STEDdepletion beam shrinks the fluorescent spot~20–50 nm
PALM / STORM (single-molecule localisation)switch few molecules on at a time and fit their centres~20 nm
Structured illumination (SIM)patterned light, image computed~100 nm (2× LM)
Super-resolution in generalbreaks the 0.2 µm diffraction limitNobel Prize Chemistry 2014 (Betzig, Hell, Moerner)
Light-sheet microscopythin sheet of light illuminates one planefast 3D imaging of cleared organs, embryos
Whole-slide scanning (digital pathology)entire slide digitised at high powervirtual microscopy, telepathology, AI analysis
Cryo-EMfrozen-hydrated molecules, image averagingnear-atomic structure; Nobel 2017 (Dubochet, Frank, Henderson)

Common fluorochromes

LabelBinds / used forEmission colour
Fluorescein (FITC)conjugated to antibodiesgreen
Rhodamine / Texas redconjugated to antibodiesred
DAPI / HoechstDNA (nuclei)blue
Acridine orangeDNA and RNAgreen (DNA) / orange-red (RNA)
Green fluorescent protein (GFP)genetically expressed tag in living cellsgreen

Other imaging and labelling techniques

TechniquePrincipleShows
Autoradiographyradioactive precursor (e.g. ³H-thymidine) incorporated; emulsion develops silver grainswhere and when molecules are synthesised
Immunohistochemistry (direct)labelled primary antibody binds antigenlocation of a protein
Immunohistochemistry (indirect)unlabelled primary, then labelled secondary antibody against itsame, with signal amplification
In situ hybridisationlabelled nucleic acid probe binds complementary DNA/RNAlocation of a gene or mRNA
Immunogold (EM)antibody bound to gold particlesantigen position at ultrastructural level

3 Traps: the correct facts for common wrong answers

✓ A 10 µm structure that appears 5 mm long in a photomicrograph is magnified about ×500.
✓ A 40x objective with a 10x eyepiece gives 400x total magnification.
✓ A basophilic component stains with haematoxylin because it is rich in acidic groups, such as nucleic acid phosphates.
✓ A cell 25 µm across measures 0.025 mm.
✓ A confocal microscope uses a pinhole to reject light from outside the focal plane.
✓ A fluorochrome emits light of a longer wavelength than the light that excites it.
✓ A new slide is first found and focused with a low-power objective.
✓ Autoradiography shows, by silver grains, where a radioactive precursor was taken up.
✓ By d = 0.61λ / NA, light-microscope resolution depends on the wavelength of the light and the numerical aperture.
✓ Cryo-electron microscopy images molecules frozen in vitreous ice, unstained.
✓ Dark structures in a transmission electron micrograph scattered many electrons because they bound more heavy metal.
✓ Fluorescein-labelled antibodies in the fluorescence microscope emit green light.
✓ Freeze-fracture electron microscopy shows the inner faces of a membrane, split along its lipid bilayer.
✓ Heavy metals in the fixative and ultrathin sections cut with a glass or diamond knife are features of transmission electron microscopy.
✓ Immersion oil raises the numerical aperture and resolution of the 100x objective.
✓ In a bright-field light microscope, the objective lens chiefly determines resolution and magnification.
✓ In dark-field microscopy, the specimen appears bright on a black background.
✓ In indirect immunohistochemistry, the visible label is carried by a secondary antibody directed against the primary antibody.
✓ In microscopy, resolution is the smallest distance at which two points are still seen as separate.
✓ Light of a shorter wavelength improves the resolution of a light microscope.
✓ Living, unstained cells are best seen with phase contrast microscopy.
✓ Magnifying beyond what the resolution supports adds no detail.
✓ Polarised light microscopy shows birefringent structures, such as collagen and crystals.
✓ Ribosomes, about 25 nm across, can be seen only with the transmission electron microscope.

4 Practise this lecture